39

8.

Select menu item [5] Note to edit

the calibration info text that is
stored in the probe. Edit the text
using the select button and arrow
keys. When done, select the OK
character in the bottom right
corner to save the changed text.

To exit without saving, press the
Cancel button.

9.

Select menu item [6] Apply to

save the calibration in the probe.
Check the applied corrections in
the confirmation screen and press
the Apply button.

To exit without applying the
correction, press the Cancel
button.