Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 39


Chemical purge
PUR [ON/OFF]

Enable or disable automatic
chemical purge

PURGE

Activate chemical purge

PURR [ON/OFF]

Enable or disable automatic
chemical purge in power up


Sensor heating
XHEAT [ON/OFF]

Enable or disable sensor heating

Measurement Output

R Start Continuous Outputting

Syntax: R<cr>

Starts output of measurements to the peripheral devices (RUN-mode); the
only command that can be used is S (stop).

The output mode can be changed with command FORM.

Example:

>r<cr>
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C

S Stop Continuous Outputting

Syntax: S<cr>

Stops the continuous output. Also the Esc key (on your computer
keyboard) can be used to stop outputting.