Modbus Communication

Addressing with Modbus Protocol

The Modbus variant used in DPT145 communication is Modbus RTU.

An example of Modbus addressing is shown below.

Request:

0x01 0x03 0x00 0x04 0x00 0x02 0x85 0xCA

addr. func.

reg. 5

2 registers CRC check

Response: 0x01 0x03 0x04 0xBC 0xC0 0x41 0xC2 0x6E 0x5E

addr. func. len.

reg. 5

reg. 6

CRC check

For a list of available Modbus registers, see Appendix A, Modbus

Reference on page 60.

Accessing Serial Port Command Interface from

Modbus Mode

You cannot change configuration settings (other than Modbus address)

while the device is in the Modbus mode. To use the configuration

commands on a Modbus-enabled device:

1.

Connect the USB service cable to the PC and install the driver, if

necessary.

2.

Open the terminal program and open a connection to the

corresponding COM port using settings 19200, 8, N, 1, no flow

control.

3.

Plug the service cable to the connector II of the instrument.

4.

In less than five seconds, send a hash character (

#

) and carriage

return (Enter) to the device using the terminal program. The device

is now ready for configuration commands. The Modbus mode is

restored on next power-up unless Modbus is disabled with

command

SMODE

.

For example, if you want to deactivate the Modbus mode on a DPT145

device, follow the procedure above and deactivate the Modbus mode

with the command SMODE STOP. On subsequent power-ups, the serial

command interface will be activated.