Mpc multipoint pressure corrections – Vaisala PTU200 User Manual

Page 39

Advertising
background image

C

HAPTER

6_______________________________________________

CALIBRATION AND

ADJUSTMENT

V

AISALA

_________________________________________________________________________ 35

MPC Multipoint pressure corrections

MPC x <cr>

where:

x = ON or OFF

The MPC command is used to activate or deactivate the multipoint
corrections. Plain command MPC outputs the corrections in use.

The multipoint corrections are protected with switch SW4 (see

F

IGURE

6-1

). Turn the switch SW4 to write ENABLE position (ON) to be able

to turn the multipoint corrections on or off.

NOTE

Changing the multipoint corrections on or off will automatically
cancel the previous date of calibration of the transmitter (see CALD
command).

Example of multipoint adjustments performed on a transmitter with
one pressure transducer:

>mpc off <cr>
Multipoint adj:

OFF

>mpc on <cr>
Multipoint adjustments

ON

>mpc<cr>

Reading Correction

Reading Correction

499.660

-0.110*******.**********.***

599.110

-0.080*******.**********.***

698.580

-0.060*******.**********.***

800.950

-0.030*******.**********.***

900.400

0.010*******.**********.***

947.200

0.020*******.**********.***

999.840

0.050*******.**********.***

1099.090

0.070*******.**********.***

Multipoint adj:

ON

>

NOTE

Always remember to return the switch SW4 to write DISABLE
position (OFF).