Alarms – Vaisala AWS330 User Manual
Page 140

User's Guide ______________________________________________________________________
138 _________________________________________________________________ M211296EN-B
Alarms
The following parameters provide limits for preconfigured alarms:
1012-168
Figure 82
Setting Alarm Limits
Table 37
Alarm Limit Parameters
Monitored Value
Unit
Alarm
Set Name
Limit Type
Default Limit
Air temperature instant
°C
High limit TA_AlarmHigh high_limit
40
Air temperature instant
°C
Low limit TA_AlarmLow Low_limit
0
Wind speed max 10 min
m/s
High limit WS_Alarm
high_limit
20
Precipitation sum 1 hour
mm
High limit PR_Alarm
high_limit
20
Datalogger temperature
°C
High limit None
Fixed +65 °C 65
Datalogger temperature
°C
Low limit None
Fixed -65 °C -65
Datalogger supply voltage VDC
Low limit DC_Alarm
Low_limit
10.5
To enable transmission of alarm message with the selected media, use
one of the options described in Table 27 on page 123.