Aalborg PWE Digital User Manual
Page 59
Advertising

54
Figure 5-1 PWE configuration jumpers
Table 5-1
Function
Analog Flow Output Analog Temp. Output* RS-485
J3A
J3B
J3C
J3D
J3E
J3F
J3G
Analog
Output
0-5 Vdc
2-3
5-6
8-9
11-12 14-15 17-18
4-20 mA
1-2
4-5
7-8
10-11 13-14 16-17
RS-485
220 Ohm
Termination
Resistor
OFF
20-21
ON
19-20
5.2
Analog Output Signals calibration
Note: The analog output available on the PWE Flow Meter was calibrated
at the factory for the specified fluid and full scale flow range (see the
device’s front label). There is no need to perform analog output
calibration unless the EEPROM IC was replaced or offset/span
adjustment is needed. Any alteration of the analog output scaling
variables in the EEPROM table will VOID calibration warranty supplied
with instrument.
,