Current measurement configuration, Ac low frequency filter – Agilent Technologies Switch/Measure User Manual

Page 50

Advertising
background image

38

34980A User’s Guide

2

Features and Functions

Current Measurement Configuration

This section contains information to help you configure the instrument
for making current measurements on the 34921A multiplexer module.
The module has four fused channels for direct dc and ac current
measurements on the ranges shown below.

Current measurements are allowed only on channels 41 through 44
on the 34921A module.

AC Low Frequency Filter

The instrument uses three different ac filters which enable you to either
optimize low- frequency accuracy or achieve faster ac settling times.
The instrument selects the slow (3 Hz), medium (20 Hz), or fast (300 Hz)
filter based on the input frequency that you specify for the selected
channels or the internal DMM.

Applies to ac current and ac voltage measurements only.

The

CONFigure

and

MEASure?

commands automatically select the 20 Hz

(medium) filter.

The instrument selects the default 20 Hz (medium) filter after a Factory

Reset (

*RST

command). An Instrument Preset (

SYSTem:PRESet

command) or

Card Reset (

SYSTem:CPON

command) does not change the setting.

Front Panel Operation:

DMM or Channel (Configure) > AC FILTER

Remote Interface Operation:

Specify the lowest frequency expected in

the input signal on the specified channels. The instrument selects the
appropriate filter based on the frequency you specify (see table above).

[SENSe:]CURRent:AC:BANDwidth {3|20|200} [,(@<ch_list>)]

If you omit the optional <ch_list> parameter, the command applies to the
internal DMM.

10 mA

100 mA

1 A

Autorange

AC low frequency filter

Input Frequency

Default Settling Delay

Minimum Settling Delay

3 Hz to 300 kHz (Slow)
20 Hz to 300 kHz (Medium)
200 Hz to 300 kHz (Fast)

7 seconds / reading
1 second / reading
0.12 seconds / reading

1.5 seconds
200 ms
20 ms