Campbell Scientific CS700 Tipping Bucket Rain Gage and CS700H Heated Rain Gage User Manual
Page 37

Appendix A. CS700H Operation
Cycle Enable
aX29!
aX29+0!
aX29+1!
– X29 Command (Only operates if “External Control” set to 1)
Get the Cycle Enable state using the aD0! Command
Set the Cycle Enable state to 0 => Heating Cycle Off
Set the Cycle Enable state to 1 => Heating Cycle On
(Changing this state forces X21 Heater Control to 0=> Off)
(This command is only available in S/W Rev 4.1 and above)
Preset / Calibrate Ambient
Temperature Sensor
aX91!
aX91+22.3!
aX91+72.1!
– X91 Command
Get the Ambient temperature sensor using the aD0! Command
Set the Ambient temperature sensor to +22.3°C (if X28 set to 0)
Set the Ambient temperature sensor to +72.1°F (if X28 set to 1)
Preset / Calibrate Funnel
Temperature Sensor
aX92!
aX92+18.6!
aX92+65.5!
– X92 Command
Get the Funnel temperature sensor using the aD0! Command
Set the Funnel temperature sensor to +18.6°C (if X28 set to 0)
Set the Funnel temperature sensor to +65.5°F (if X28 set to 1)
Set Factory Defaults
aX99+1!
– X99 Command
Set the factory defaults (only when data is +1)
A-9