Measurement instruction, 4. wiring for example 2 – Campbell Scientific HFP01SC Hukseflux Self-Calibrating Soil Heat Flux Plate User Manual
Page 15

Model HFP01SC Self-Calibrating Soil Heat Flux Plate
6.3.2 Example 2. Sample CR10(X) Program Using a Single-Ended
Measurement Instruction
TABLE 6-4 provides the wiring for Example 2.
TABLE 6-4. Wiring for Example 2
Description
Color
CR10(X)
Sensor Signal
White
1H
Sensor Signal Reference
Green
AG
Shield
Clear
G
Heater Resistor Signal
Yellow
1L
Heater Resistor Signal
Reference
Purple
AG
Shield
Clear
G
Power
Red
SW12
Power Reference
Black
G
External Power Control
jumper wire SW12-CTRL to
C8
;{CR10X}
;
*Table 1 Program
01: 1
Execution Interval (seconds)
;Measure HFP01SC on smaller range.
;
1: Volt (SE) (P1)
1: 1
Reps
2: 22
7.5 mV 60 Hz Rejection Range
3: 1
SE Channel
4: 2
Loc [ shf_mV ]
5: 1
Mult
6: 0
Offset
;Measure HFP01SC on larger range.
;
2: Volt (SE) (P1)
1: 1
Reps
2: 23
25 mV 60 Hz Rejection Range
3: 1
SE Channel
4: 8
Loc [ shf_mV_a ]
5: 1
Mult
6: 0
Offset
;Load in the factory calibration.
;
3: If (X<=>F) (P89)
1: 3
X Loc [ cal ]
2: 1
=
3: 0
F
4: 30
Then Do
9