Model HFP01SC Self-Calibrating Soil Heat Flux Plate

6.3.2 Example 2. Sample CR10(X) Program Using a Single-Ended

Measurement Instruction

TABLE 6-4 provides the wiring for Example 2.

TABLE 6-4. Wiring for Example 2

Description

Color

CR10(X)

Sensor Signal

White

1H

Sensor Signal Reference

Green

AG

Shield

Clear

G

Heater Resistor Signal

Yellow

1L

Heater Resistor Signal

Reference

Purple

AG

Shield

Clear

G

Power

Red

SW12

Power Reference

Black

G

External Power Control

jumper wire SW12-CTRL to

C8

;{CR10X}
;
*Table 1 Program
01: 1

Execution Interval (seconds)


;Measure HFP01SC on smaller range.
;
1: Volt (SE) (P1)
1: 1

Reps

2: 22

7.5 mV 60 Hz Rejection Range

3: 1

SE Channel

4: 2

Loc [ shf_mV ]

5: 1

Mult

6: 0

Offset


;Measure HFP01SC on larger range.
;
2: Volt (SE) (P1)
1: 1

Reps

2: 23

25 mV 60 Hz Rejection Range

3: 1

SE Channel

4: 8

Loc [ shf_mV_a ]

5: 1

Mult

6: 0

Offset


;Load in the factory calibration.
;
3: If (X<=>F) (P89)
1: 3

X Loc [ cal ]

2: 1

=

3: 0

F

4: 30

Then Do

9