3 fft window, 4 histogram bin width, 5 data to collect – Cirrus Logic CDB5467U User Manual

Page 21: 6 cycle count, 7 accept button, 8 collecting data sets, 9 retrieving saved data from a file, Cdb5467u

Advertising
background image

CDB5467U

DS714DB1

21

2.8.7.3

FFT Window

This box allows the user to select the type of windowing algorithm for FFT processing. Windowing algo-
rithms include the Blackman, Blackman-Harris, Hanning, 5-term Hodie, and 7-term Hodie. The 5-term
Hodie and 7-term Hodie are windowing algorithms developed at Crystal Semiconductor.

2.8.7.4

Histogram Bin Width

This field determines the "bin width" when plotting histograms of the collected data. Each vertical bar in
the histogram plot will contain the number of output codes entered in this field. Increasing this number
may allow the user to view histograms with larger input ranges.

2.8.7.5

Data to Collect

These two check boxes allow the user to select the data types that will be collected and returned to the
PC for processing.

2.8.7.6

Cycle Count

The value in the Cycle Count field will be written to the Cycle Count register in the CS5467. The Cycle
Count register determines the length of one computation cycle. The Cycle Count value should be selected
appropriately according to the Data to Collect setting. For example, if the Data to Collect is instantaneous
current, voltage, or power it is better to set Cycle Count to 1.

2.8.7.7

Accept Button

When this button is pressed, the current settings will be saved, and the user will return to the Data Col-
lection Window
.

2.8.8

Collecting Data Sets

To collect a sample data set:

1. In the Data Collection Window, press the Config button to bring up the Configuration window and view the

current settings.

2. Select the appropriate settings from the available options (see the Configuration Window section) and press

the Accept button.

3. The Data Collection Window should still be visible. Press the Collect button to begin collecting data.

4. Once the data has been collected, it can be analyzed, printed, or saved to disk.

2.8.9

Retrieving Saved Data From a File

The CDB5467U software allows the user to save data to a file, and retrieve it later when needed. To load
a previously saved file:

1. Pull down the Setup menu and select the Disk menu item. A file menu will appear.

2. Find the data file in the list and select it. Press the Select button to return.

3. Go to the Data Collection Window, and press the Collect button.

4. The data from the file should appear on the screen. The data will be ready for different types of analysis.

5. To select a different file, repeat the procedure.