RS232 Serial Port

The MMX-6 is equipped with an RS232 serial port. Using the

accessory cable (part# N-306-0010), the MMX-6 has the ability to connect

to a computer, or external storage device. The following section outlines

the procedure for connecting the MMX-6 to a computer, and how to collect

data using any standard communications program:

Connecting To a Computer

1) Connect the accessory cable (part# N-306-0010) to the 2 pin jack

located on the bottom of the MMX-6, and the 9 pin connector to a

serial port on the computer.

2) Start the communications software that will be used to collect the

measurements (i.e. Microsoft Windows

3.1 -Terminal, or 95, 98,

Me, XP - HyperTerminal).

3) Setup the communications software using the following

parameters: Data Bits - 8, Parity - None, Stop Bits - 1, Baud

Rate 1200.

4) Set the communications software COMM port to the port number

that the MMX-6 is connected.

5) After taking a measurement, press the SEND key to send the

measurement to the computer. The measurement will be displayed

on the computer screen.