INFICON Helium Leak Detector Modul1000 User Manual

Page 65

Advertising
background image

Operation

6-24

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

RS232

The Modul1000 is controlled via RS232 interface by an external computer. In this
mode, the Modul1000 cannot be controlled via the keyboard.

All

Control interfaces PLC, RS232 and Local

Local and PLC

The Modul1000 is controlled via the START, STOP and ZERO buttons on the device
and via the digital inputs.

Local and RS232

The Modul1000 is controlled via the START, STOP and ZERO buttons on the device
and via the digital inputs.

Local

The Modul1000 is controlled via the START, STOP and ZEROR buttons.

Main menu

Settings

Interfaces

RS232

Local

Without request, the leak detector continuously sends the status and the leak rate.

Binary

Allows reading of equipment parameters, e.g. during maintenance.

UL2xxLeak Ware

Allows controlling and reading of measured values when connected to a computer.
(For operation of the Leak Ware refer to the corresponding operating instructions).

Notice:

The calibration function of the Leak Ware cannot be used for the
Modul1000. Please execute the function „STORE DATE“ in the operating
mode „Single Part Measurement“ for recording the measured values.

ASCII

Enables the operation of the Modul1000 via an RS232 terminal. For further details
refer to Interface Description.