9 rs-232/lan connector, 10 motor connector – INFICON CI-100 Crucible Indexer User Manual
Page 26
2 - 10
IP
N 07
4-
54
4-
P1
B
CI-100 Crucible Indexer Operating Manual
2.9 RS-232/LAN Connector
A DB-9 to RJ-45 adapter and standard Ethernet cable are supplied for connecting
the CI-100 to the PC for RS-232 communications.
NOTE: CI-100 does not support Ethernet communications.
2.10 Motor Connector
The controller cable supplied is a DB-25 male to female straight-through 26 AWG,
shielded. The standard length is 10 ft. (3 m). For longer lengths, use larger gauge
wire to avoid power loss to the motor drive.
Table 2-3 RS-232/LAN Connector
RJ-45 Pin
Signal
DB-9F Pin
1 (Wht/Org)
Unused
2 (Org)
Unused
3 (Wht/Grn)
Unused
4 (Blu)
RS-232 Tx
2 (Red)
5 (Wht/Blu)
RS-232 Rx
3 (Grn)
6 (Grn)
Unused
7 (Wht/Brn)
RS-232 Ground
5 (Brn)
8 (Brn)
Unused
Table 2-4 Motor Connector
Pin
Signal
1
DAC CSp
2
DAC CLKp
3
DAC DATAp
4
STEPp
5
DIRp
6
Encoder CLKp
7
Encoder CSp
8
Encoder DATAp
9
Ground
10
Ground
11
Ground
12
24 V (dc)
13
24 V (dc)
14
DAC CSn