JULABO FC600-T Recirculating cooler User Manual

Page 43

Advertising
background image

Recirculating coolers

43


ACTUAL

SET

Ad¯P -

Calibration of programmer input:

Highest value = 80 °C

ACTUAL

SET

AdP -

Calibration of programmer input:

Lowest value = -20 °C

ACTUAL

SET

Calibration procedure:

In the configuration level, set the programmer input

type to allow control via an external programmer.
0 = Voltage 0 to 10 V
1 = Current 0 to 24 mA

Connect an external programmer to the connector (O).

Set the external programmer to the highest

temperature value.

ACTUAL

SET

In the calibration level, select the item Ad¯P and set

the highest temperature value
(example: 80 °C).

Press the setpoint key (E).

Follow the instructions under section 6.3.1.
page 27




Do not alter any settings on either of the units as long

as the display "Ad¯P" is blinking.