Page 88

•
Frequency — 1 Hz
•
Wave Shape — Ramp
•
Compensator — None
Important: It is important to have Compensator set to None. If it is left on, it can negate
the function generator signal.
5. Open a Station Manager Scope.
Click the Scope button in the Station Manager main window.
6. Set up the Scope.
Set up a Scope to compare the command signal to the resulting feedback. In the Station Manager main
window, click the Signal Scope button. Enlarge the scope if necessary, and then set up the scope as
follows:
Y1 parameters:
•
Channel — Ch1 (or other depending on your configuration)
•
Signal — Command
•
Unit — kN
•
Units/div — 0.5
•
Offset — 2
Y2 parameters:
•
Channel — Ch1 (or other depending on your configuration)
•
Signal — Force
Note: Another possible selection is Active Feedback. Active Feedback will automatically
plot your active channel (for example force or displacement).
•
Unit — kN
88 | MTS Landmark Test System Operation
Running the Example Tension Test