Ml/ h r – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 26

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

21

09/18/14

Phase

Function

5

LOP 02

Phase

Function

Rate

Volume

Direction

6

RAT

750.0 mL/hr

0.5 mL

Infuse

Phase

Function

Rate

Volume

Direction

7

RAT

300.0 mL/hr

1.5 mL

Infuse

Phase

Function

8

BEP

Phase

Function

9

LPS

Phase

Function

10

PAS 60

Phase

Function

Rate

Volume

Direction

11

RAT

500.0 mL/hr

3.75 mL

Infuse

Phase

Function

12

LOP 03

Phase

Function

Rate

Volume

Direction

13

RAT

900.0 mL/hr

17.25 mL Withdraw

Phase

Function

14

BEP

Phase

Function

15

PAS 00

Phase

Function

16

LPE

7.4.5

Example 5: Control from a High-Low Pressure Sensor

10

25

m

l/

h

r

Pumping
Rate

Program
Output Pin

Event Input
Pin

Pressure high sensor

Pressure low sensor

Pressure trigger points

This example demonstrates a Pumping Program whose control depends on an external sensor. Assuming a
pressure sensor that is configured to detect a high pressure point and a low pressure point, the Pumping
Program individually selects whether it will react to the high or low pressure point.

The “Program Output” pin on the TTL I/O connector (pin 5) is used to select the high or low pressure point.
When low, the low pressure point is selected (PHASE 01), and when high, the high pressure point is selected
(PHASE 05). The Program begins by infusing continuously at 10.0 mL/hr (PHASE 02), while a background
trap is set for the low pressure point (PHASE 03). To create a delay when the pressure sensor is switched from
high pressure to low pressure when the “Program Output” pin is set, a small volume is pumped (PHASE 02,
06) before the background traps are set.

Advertising