35

08/01/14

10.2.5

Network Command Burst

The Network Command Burst feature is only applicable when communicating to a network of pumps.

This special feature allows commands to be sent to a network of pumps simultaneously. For example,
changing the pumping rates simultaneously on a network of pumps.

Note: Since this special feature violates the general communications protocol of one command-one response,
all of the pumps will be responding simultaneously, and therefore the communications response to a Network
Command Burst will be gibberish and should be ignored.

Command Format
Command Burst => <n> <command> *
Network Command Burst => < Command Burst > [Network Command Burst]

Where <n> indicates the address of the pump that is to execute <command>. Maximum address is 9.
n => {0 ... 9}

Example: Simultaneously change the pumping rates of 3 pumps on a pump network as follows:

Pump 0: 100 mL/hr
Pump 1: 250 mL/hr
Pump 2: 375 mL/hr


Assuming that the current pumping rate units of all 3 pumps are currently mL/hr, send the following
command, followed by a carriage return <CR>, (spaces are optional):

0 rat 100 * 1 rat 250 * 2 rat 375 *

10.3

Command Errors and Alarms

If a command received by the pump is not recognized or the data is invalid, an error message will be in the
<data> field of the response packet following the <prompt> field. The following are the error responses:

<command error> => ? <error>

<error> =>

λ

Command is not recognized (‘?’ only)

NA

Command is not currently applicable

OOR

Command data is out of range

COM

Invalid communications packet received

IGN

Command ignored due to a simultaneous new Phase start

When an alarm occurs, the alarm must be acknowledged before any data is changed or the pump is started.
Alarms are acknowledged by the user clearing the alarm message on the keypad, or the alarm status being sent
in response to any valid RS-232 command. An alarm message sent automatically in the Safe Mode will not
clear the alarm condition. This is to verify that the alarm message was sent to a receptive host, such as after a
power failure when both the computer and the pump were reset. In this case, the pump will most likely send
its reset alarm message before the computer has finished booting.

10.4

RS-232 Command Set

All data changed from RS-232 is stored in the non-volatile memory, except for changes to the pumping rate
while pumping. All “Program Phase Data” refers to the currently selected Program Phase. Use the Phase
select command (‘PHN’) to query or select the current Phase. A Phase consists of the pumping rate, ‘Volume
to be Dispensed’, and the pumping direction.

A packet without a command is interpreted as a status query. The addressed pump responds with a status only
response packet.

Except where noted, a command without any parameters is a query command. The response packet data will
include the requested data. In general, the query response data will be in the same format as the parameters for
setting the command. For example, the query diameter command ‘DIA’ will respond with ‘<float>‘ as the
response ‘<data>‘.

Otherwise, the command is a set command. If the data was set, a status only response packet will be sent. If
the data was not set, the response packet will include an error (<command error>) message indicating why the
data was not set.