Conditions for the place of installation
Ambient temperature:
5°C to 40°C

If the ambient temperature drops below or exceeds

these values the electronic and mechanical components

can become damaged and performance data are

changed to an unknown extent.

Atmospheric humidity:
Maximum relative humidity 80% at temperatures up

to 31°C,
decreasing linearly down to 50% relative humidity at

40°C.

At higher atmospheric humidity the electronic and

mechanical components can become damaged and

performance data are changed to an unknown extent.

Installation height:

max. 2000 m above sea level

Installation

Mount the MM 400 on a firm, stable laboratory bench

since otherwise vibrations can be transmitted.

Remove transport safeguard:
A transport safeguard, which is marked with an arrow,

is located on the underside of the MM 400.

Unscrew the screw and remove.

Keep the transport safeguard device for transport at

a later date.

Operation with the transport safeguard or transport

without the transport safeguard can cause damage to

the mechanical components.

Electrical connection

The voltage and frequency for the MM 400 are

given on the type plate.

Ensure that these values correspond to the

available power supply system.

Connect the MM 400 to the power supply system

using the supplied connection cable.