Sunnex PF Series MRI Service & Installation Manual

24

Guidance and Manufacturer’s Declaration—Immunity All Equipment and

Systems

The PF MRI Light (PF 28016-M) is intended for use in the electromagnetic
environment specified below. The customer or user of the PF MRI Light (PF 28016-M)
should assure that it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance
Level

Electromagnetic Environment -
Guidance

Electrostatic
Discharge (ESD)
IEC 61000-4-2

±6kV contact

±8kV air

±6kV contact

±8kV air

Floors should be wood, concrete
or ceramic tile. If floors are
synthetic, the relative humidity
should be at least 30%.

Electrical Fast
Transient/burst
IEC 61000-4-4

±2kV on AC
Mains

±2kV on AC
Mains

Mains power quality should be that
of a typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV
Differential
±2kv Common

±1 kV
Differential
±2kV
Common

Mains power quality should be that
of a typical commercial or hospital
environment.

Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11

>95% Dip for
0.5 Cycle

60% Dip for
5 Cycles

30% Dip for
25 Cycles

>95% Dip for
5 Seconds

>95% Dip for
0.5 Cycle

60% Dip for
5 Cycles

30% Dip for
25 Cycles

>95% Dip for
5 Seconds

Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the PF
MRI light requires continued
operation during power mains
interruptions, it is recommended
that the PF MRI Light (PF 28016-
M) be powered from an
uninterruptible power supply or
battery.

Power
Frequency
50/60Hz
Magnetic Field
IEC 61 000-4-8

3A/m

3A/m

Power frequency magnetic fields
should be that of a typical location
in a typical commercial or hospital
environment.