22

Incidental function

Usage of external output

attention :

The program method, the set method, and the communication interface are

described in detail to the appended owner guide only for the controller. Refer to it.

1. Specification of connector

From the back of Thermo Elite

Pin number

Name of signal

DC+12V

Time-up output(open collector)

DC+12V

External warning output (open collector)

Temperature output (+)

Temperature output (-)

Shield (analog signal)

2. Notes concerning liquid used

Unit of output voltage : 5mV/

The output is 0mV at 0

℃ in measurement temperature.

Output

voltage(V)

Measurement temperature(

℃)

3. External warning output

Transistor open collector output

Output load

Power supply DC12V

sink current

MAX 100mA

Specificatio
n of out-put
relay

MATSUSHITA JAI-F-TM-
DC12V / OMRON G2R-1A-T
DC12V

TROUBLE
sign (Blink
to a main
display the
err-or code)

External
warning
output

4. Time-out output

Transistor open collector output

Output load

Power supply DC12V

sink current

MAX 100mA

Specification
of out-put
relay

MATSUSHITA JAI-F-TM-
DC12V / OMRON G2R-1A-T
DC12V

Blink to the
FI-NISH
indicator

Time-out
output

Blinking

Blinking

ON

ON

Advertising