7 technical specifications, 1 reference conditions, 2 amperometric detector – Metrohm 896 Professional Detector – Amperometry User Manual
Page 48: Reference conditions, Amperometric detector

7.1 Reference conditions
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
2.896.0020 896 Professional Detector – Amperometry
7 Technical specifications
7.1
Reference conditions
The technical data listed in this Chapter refers to the following reference
conditions:
Ambient tempera-
ture
+25 °C (± 3 °C)
Instrument status
> 40 minutes in operation (equilibrated)
7.2
Amperometric detector
Type
Microprocessor-controlled Digital Signal Processing (DSP technology)
Potentiostat
Potential range
–5.0…+5.0 V in steps of 0.001 V
Potential step
response time
< 1 ms
Detection
modes
■
DC
■
PAD
■
flexIPAD (flexible IPAD)
■
CV
Measuring unit
AutoRange
yes, (DC only)
Digital signal
range
DC mode
0.00012 pA…2 mA
PAD mode
0.012 pA…2 mA
flexIPAD mode
0.12 pC…200 µC
CV
0.12 pA…20 mA
Electronic noise
DC mode
< 5 pA
PAD mode
< 10 pA
flexIPAD mode
< 30 pC
Filter
DC mode
Hardware filter, can be selected by the user
All modes
Software filter, can be set by the user