3 polarizer, Polarizer – Metrohm 905 Titrando User Manual
Page 46

8.1 Measuring interface
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
905 Titrando
Resolution
Pt1000
0.1 °C
NTC
0.1 °C
Measuring accu-
racy
Pt1000
±0.2 °C
(Applies for measuring range –20 - +150 °C; ±1 digit; without sensor
error, under reference conditions)
NTC
±0.6 °C
(Applies for measuring range +10 - +40 °C; ±1 digit; without sensor
error, under reference conditions)
8.1.3
Polarizer
One measuring input (Pol.) for polarizable electrodes.
Measuring mode
Ipol
Determination with adjustable polarization current
Polarization
current
–122.5 - +122.5 µA (increment: 0.5 µA)
–125.0 - +125.0 µA: non-guaranteed values, dependent on reference
voltage +2.5 V
Measuring
range
–1,200 - +1,200 mV
Resolution
0.1 mV
Measuring
accuracy
±0.2 mV
(±1 digit, without sensor error, under reference conditions)
Measuring mode
Upol
Determination with adjustable polarization voltage
Polarization
voltage
–1,225 - +1,225 mV (increment: 25 mV)
–1,250 - +1,250 mV: non-guaranteed values, dependent on reference
voltage +2.5 V
Measuring
range
–120 - +120 µA
Resolution
0.1 µA