Figure 8, Installing tubing weighting and aspiration filter – Metrohm 940 Professional IC Vario ONE/ChS/HPG User Manual

Page 31

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

940 Professional IC Vario ONE/ChS/HPG (2.940.1240)

■■■■■■■■

23

2 Mounting the tubing adapter and the aspiration filter

Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:

Start by pushing the tubing weighting onto the loose end of the
eluent aspiration tubing.

Then push the clamping screw onto the loose end of the eluent
aspiration tubing.

Lastly, push the filter holder onto the loose end of the eluent aspi-
ration tubing and screw it onto the tubing nipple.
The end of the tubing should extend approximately 1 cm.

Installing the aspiration filter:

NOTE

Only handle the aspiration filter while wearing gloves.

Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.

Tighten the aspiration filter to the filter holder.

Figure 8

Installing tubing weighting and aspiration filter

3 Installing the eluent bottle cap on the eluent bottle

Insert the eluent aspiration tubing into the eluent bottle
(6.1608.070).

Tighten the bottle cap on the eluent bottle.

Adjust the length of the eluent aspiration tubing so that the aspi-
ration filter is at the bottom of the eluent bottle. Then fasten it in
place using the M8 tubing nipple.

Seal the M6 opening on the bottle cap with the M6 threaded
stopper from the accessory set.

Advertising