3 troubleshooting, 1 problems, Troubleshooting – Metrohm 835 Titrando User Manual

Page 39: Problems

Advertising
background image

3

Troubleshooting

Titrando Installation Instructions

33

3 Troubleshooting

3.1 Problems

Problem

Possible cause

Measures

"On" LED does not

light up although the

Titrando is con-

nected to the mains

supply.

Either the Touch Control or

the computer is not switched

on, or the plugs are not

plugged in correctly.

Check the plug connections and

switch on the Touch Control or the

computer.

Titrando with internal

dosing drive only:

"Status" LED does

not light up although

an exchange unit is

in place.

The exchange unit has not

been attached correctly.

Remove the exchange unit and

push it on again until it clicks into

position. The LED blinks while the

data is being read out from an intel-

ligent exchange unit (806) and is on

continuously when the exchange

unit has been recognized correctly.

The exchange unit flat cock

is not in the change position.

Move the flat cock manually to the

change position (switch lever point-

ing to the right).

Titrando with internal

dosing drive only:

The exchange unit

cannot be pushed

into position.

The piston rod in the ex-

change unit is not in the cor-

rect position.

Move the piston rod to the correct

position (see exchange unit Instruc-

tions for Use, Section "Attaching

the exchange unit").

Titrando with internal

dosing drive only:

The exchange unit

cannot be removed

and the "Status" LED

blinks slowly.

Dosing or filling is currently

taking place and / or the Ti-

trando is not in the change

position.

Stop the run or carry out a "Fill"

step.