4 Notes - Maintenance - Faults
819 IC Detector / 820 IC Separation Center
76
6 Insert connection piece
•
Insert connection piece 73 in suppressor holder 75 in such a
way that connection "1" is at the top, and that the three lugs
on the connection piece fit in the corresponding openings of
the holder.
•
Screw nut 72 onto the thread of suppressor holder 75 manu-
ally (do not use tools).
7 Connect and condition the suppressor
•
Reconnect the suppressor to the IC system.
•
Before switching the suppressor to the next position for the
first time, rinse all 3 suppressor units with solution for 5 min.
Advertising