Figure 11, Eluent aspiration tubing fully equipped – Metrohm 882 Compact IC plus – Cation User Manual
Page 31
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
882 Compact IC plus – Cation
■■■■■■■■
23
■
Slide the tubing weighting (10-3) onto the eluent aspiration tub-
ing (10-1).
■
Slide the clamping screw (10-4) onto the eluent aspiration tubing
(10-1).
■
Insert eluent aspiration tubing (10-1) into the aspiration filter
(10-5). The end of the tubing should approximately reach to the
center of the aspiration filter.
■
Screw together clamping screw (10-4) and filter holder (9-1).
Figure 11
Eluent aspiration tubing fully equipped.
5 Mounting eluent aspiration tubing to the eluent bottle
■
Insert the eluent aspiration tubing into the eluent bottle (12-10).
■
Fasten the completely equipped bottle attachment (10-2) on the
eluent bottle (12-10). The aspiration filter (12-6) must rest on the
base of the eluent bottle.
■
Close the remaining small opening on the bottle attachment with
a threaded stopper (12-14) from the accessories set.
6 Mounting the adsorber tube
Note
If alkaline eluents and eluents with lower buffer capacity are used,
the eluent bottle must be equipped with an adsorber tube filled
with CO
2
■
First, place a piece of cotton (12-3), then the CO
2
adsorber (12-4)
in the large opening of the adsorber tube (12-2) and close with
the plastic cover.
■
Fasten the adsorber tube (12-2) onto the bottle attachment
(12-11) using the SGJ clip (12-12).