2.8 Dosing and liquid handling

Metrohm 838 Advanced Sample Processor, Operation

57

The contents of the dosing cylinder is ejected via the selected port. The

standard PREP port (Port *) is that defined under
>dosing drive X

preparation port Y
(default: Port 1).

Empty

DOS: X.Y : empty mL

Completely empties the dosing cylinder

and tubing

The tubing system and Dosing unit cylinder can be completely emptied.

The liquid in the dosing cylinder is ejected via the dosing port. The air

required to displace the liquid from the tubing is aspirated via the given

port. The standard port (Port *) for aspirating the air is that defined un-

der
>dosing drive

X

drain port Y
(default: Port 4).
The standard dosing port can be altered under
>dosing drive X

dosing port Y
(default: Port 1).

Exchange Dosing unit

DOS: X.Y : release mL

Prepares Dosino for exchanging the Dosing

unit.

Before changing the Dosing unit the

release

command must be used

to fill the dosing cylinder and move the stopcock to the exchange posi-

tion. The cylinder is filled by aspirating the necessary volume via the

given port. If * is entered as a wild card then the port given under
>dosing drive X

rinsing port Y
will be used (default: Port 2).