Input range
0 - 5
Default value
2
Result unit
The result unit is displayed and saved together with the result.
Selection
% | ppm | mg/mL | g | mg | mL | mg/piece |
User-defined
Default value
%
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
7.1.9
Statistics
Menu
▶ Parameters ▶ Statistics
The statistics calculation of a multiple determination is switched on under
Statistics and it is defined how many determinations the series contains.
Statistics
If this funtion is switched on, then statistics calculations will be carried out
for all of the defined results.
Selection
on | off
Default value
off
For the methods "Titer Ipol" and "Titer Upol" the
default value is on.
Number of samples
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination must be added to the determination series,
because one determination was incorrect, for example, then this can be
accomplished in the statistical overview (see Chapter 5.8, page 30).
Input range
2 - 20
Default value
3
7.1.10
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.