Figure 16, Connecting the iconnect, Figure 17 – Metrohm 904 Titrando User Manual
Page 34: Connecting an electrode to the iconnect
3.5 Connecting sensors
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
904 Titrando
Take care to ensure that the marking on the plug matches the mark-
ing on the Titrando as shown in the figure.
Figure 16
Connecting the iConnect
The iConnect is detected automatically and entered as measuring
input into the device properties of the Titrando. If an electrode is
connected to the iConnect that is not yet included in the list of sen-
sors for the control software, then a corresponding message will be
displayed.
The iConnect can be plugged in and unplugged while the Titrando is
switched on.
An electrode with integrated data chip, referred to as iTrode, is connected
to the iConnect.
Connect the electrode as follows:
1 Remove the protective caps on the iConnect and the electrode.
2 Plug in the electrode on the iConnect as illustrated below.
Figure 17
Connecting an electrode to the iConnect
The guide pin guarantees correct connection in such a way that the
contact pins cannot be damaged.
3 Screw the iConnect tightly.