Figure 19, Deaerating the high pressure pump – Metrohm 850 Professional IC Cation - Gradient User Manual
Page 44

3.10 High pressure pump
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
850 Professional IC Cation – HP-Gradient
5
1
2
6
7
5
3
4
5
Figure 19
Deaerating the high pressure pump
1
Syringe 10 mL 6.2816.020
For aspirating the eluent.
2
Luer connector
On purging needle.
3
Purging needle 6.2816.040
4
Deaerating capillary
5
PEEK pressure screws, short
6.2744.070
6
Purge valve
7
Purge valve rotary knob
1 Connecting the purging needle
■
Push the end of the purging needle (19-3) over the end of the
deaerating capillary (19-4) on the purge valve.
2 Connecting the syringe
■
Insert syringe (19-1) in the Luer connector (19-2) of the purging
needle (see Figure 19, page 36).
3 Opening purge valve
■
Open the rotary knob (19-7) by approx.
½ rotation counterclock-
wise.
4 Setting the flow rate
■
Start MagIC Net (if not yet started).
■
Ensure that the eluent aspiration tubing is immersed sufficiently in
the eluent.
■
Let the high pressure pump run.