Metrohm 756 KF Coulometer User Manual
Page 70
4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
62
│ │ ├
.Cell
Type of generator electrode
no diaph.,diaphragm 4.2.2.19.
│ │ ├
.GenI
Switching of generator I
100, 200, 400, auto ditto
│ │ ├
.Oven
KF Oven connected
COM1, COM2, no 4.2.2.20.
│ │ ├
.ActPulse
Output of a pulse
first, all, cond., OFF 4.2.2.21.
├
.Def
Definitions for data output
│ ├
.Formulas
Calculation formulas
│ │
├
.1
for result 1
│ │
│
├
.Formula Calculation
formula
special
4.2.2.22.
│ │
│
├
.TextRS
Text for result output
up to 8 ASCII char
ditto
│ │
│
├
.Decimal
Number of decimal places
0...2...5 ditto
│ │
│
├
.Unit
Unit for result output
up to 6 ASCII char
ditto
│ │
│
├
.Limits
Limits for result
ON, OFF ditto
│ │
│
├
.LoLim Lower
limit
0...
±
999 999
ditto
│ │
│
├
.UpLim Upper
limit
0...
±
999 999
ditto
│ │
│
├
.Output
Output on L13
active, pulse, OFF ditto
│ │
│
:
up to 9 results
│ ├
.SiloCalc
Silo calculations
│ │
├
.Assign Assignment
│ │
│
├
.C24
Store as variable C24
RSX,H2O,CXX
4.2.2.23.
│ │
│
├
.C25
Store as variable C25
RSX,H2O,CXX
│ │
├
.MatchId
Matching of Id's
id1, id1&2, all, OFF
│ ├
.ComVar
Assignment of common variables
│ │
├
.C30 for
C30
RSX,H2O,CXX,MNX
4.2.2.24.
│ │
├
up to C39
│ ├
.Report
Reports at the end of determination
│ │
├
.Internal
Output to internal printer (only 756) special
4.2.2.25.
│ │
├
.Assign1
Output to COM1
special
│ │
├
.Assign2
Output to COM 2
as COM1
│ ├
.Mean
Assignment for mean calculation
│ │
├
.1 MN1
│ │
│
├
.Assign
Input of variable
RSX, H2O, CXX
4.2.2.26.
│ │ │
:
├
.CFmla
Calculation constants
│
├
.1
Calculation constant C01
│ │
├
.Value
Input of value
0...±999 999
4.2.2.27.
│
├