5.8 Servicing the high-pressure pump
■■■■■■■■■■■■■■■■■■■■■■
70
■■■■■■■■
930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)
NOTE
The maximum treatment time in an ultrasonic bath is 20 s. Ultra-
sonic baths that last longer than this can damage the ruby ball
inside the valve.
The outlet valve has to be replaced if it remains clogged after the
ultrasonic bath.
3 Reinserting the outlet valve into the pump head
■
Insert the outlet valve into the outlet valve holder (the seal must
be visible) (1).
■
Screw the outlet valve holder up into the pump head (2) and
tighten it in place using the adjustable wrench (3).
■
Tighten the connection capillary to the auxiliary piston back onto
the outlet valve holder.
Servicing the inlet valve
Accessories
For this step you need the following accessories:
You can find these parts in the accessory kit: Vario/Flex Basic
(6.5000.000).
■
Adjustable wrench (6.2621.000) from the accessory kit: Vario/Flex
Basic (6.5000.000)