3 mounting the sample insert, Mounting the sample insert, Figure 5 – Metrohm 874 USB Oven Sample Processor User Manual

Page 21

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

874 Oven Sample Processor

■■■■■■■■

13

3.3

Mounting the sample insert

The dimensions of the supplied sample insert are optimized for the usage
of 6.2419.007 sample vials (6 mL) from Metrohm . This ensures the
best possible transfer of heat between oven and sample.

1

2

3

6.2063.020

6.2627.010

Figure 5

Mounting the sample insert

Note

The sample insert may not be inserted or removed unless it has been
cooled down.

Proceed as follows:

1 Insert the 6.2627.010 oven insert with the spring in place into the

oven from above and slide it all the way to the bottom.

2 Place the 6.2063.020 sample insert into the oven from above.

3 Tighten the sample insert with the supplied hexagon screws. The

required hexagon key is part of the accessories.