3 "lamp info" dialog, 4 "lamp setup" dialog, Lamp info" dialog – Metrohm 909 UV Digester User Manual
Page 29: Lamp setup" dialog

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
909 UV Digester
■■■■■■■■
21
■
The sample setpoint temperature (Sample Temp.).
■
The radiating time (Rad. Time).
Sample Temp.
Setpoint temperature for the samples.
Input range
80 - 100 °C (Increment: 1)
Default value
90 °C
Rad. Time
Radiating time for the samples.
Input range
1 - 999 min (Increment: 1)
Default value
90 min
Note
■
The configured parameters can be modified during digestion.
■
The configured parameters remain stored in the instrument after it is
switched off.
4.2.3
"Lamp info" dialog
You can open the Lamp info dialog from the main dialog by pressing the
[OK] key twice.
The following information is displayed:
■
The cumulative number of operating hours for the current UV lamp in
use since the last replacement (Hours).
■
The number of times the current UV lamp in use has been switched on
since the last replacement (Ignitions).
4.2.4
"Lamp setup" dialog