2 start/stop hardware and record baseline – Metrohm IC Net 2.2 User Manual
Page 36
4 Systems
IC Net 2.2
28
SYSTEM / Control / Disconnect from workplace
Disconnect selected system from the workplace. If a system is
disconnected, manual control is not available for this system, but
all other system settings can be modified and saved.
Start deter-
mination
and
Startup hardware
is only possible if no other system
from the same system folder is connected. In this case, the sys-
tem is automatically connected.
4.3.2
Start/stop hardware and record baseline
SYSTEM / Control / Startup hardware (Measure Baseline)
With this menu item the following actions are executed automati-
cally:
•
Send
System startup values
to the devices.
•
Start Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
•
Start Metrohm additional modules and Bischoff additional
modules.
•
Start Bischoff detectors.
•
Switch Metrohm Autosamplers and Bischoff Autosamplers to
remote control.
•
Start recording of the measurement signal using the method of
the connected system. Independently of the set chromatogram
Duration
, the measurement signal is recorded until the data ac-
quisition is stopped if a new determination is started with
Start
determination
or by clicking the icon of the chromatogram
window. In this case the user is asked if the recorded baseline
should be saved or not.
SYSTEM / Control / Shutdown hardware
With this menu item the devices are stopped as follows:
•
Stop Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
•
Stop Metrohm additional modules and Bischoff additional
modulesVariotherm.
•
Stop Bischoff detectors.
•
Switch Metrohm Autosamplers and Bischoff Autosamplers to
local control.
•
Stop running determination.
•
Stop active sample queue.
SYSTEM / Control / Stop data acquisition
Stop recording of the baseline.