Metrohm 817 Bioscan User Manual
Page 54

3 Operation
817 Bioscan
46
List of program instructions
The following program instructions can be added to the time
program on the
Program
subpage:
Instruction Parameter
entry Meaning
Mode
DC, Pulse, Scan
Select detector mode.
Cell On,
Off
Switch cell on/off.
Outputs
0, 1, *
Set remote lines Aux1, Aux2, Cont1
and Cont2 to desired values. For entry
of the first value, enter
1
,
0
or
*
. For
entry of the other values, move the
cursor in front of the value to be
changed and enter
1
,
0
or
*
.
Range
pA: 10 pA – 5 nA
nA: 0.1 nA – 50 nA
µA: 10 nA – 5 µA
Set measuring range.
Offset
-50, -40, -30,…50 %
Set offset for measuring signal.
Valve Fill,
Inject
Switch connected injection valve.
Oven
off, 15 ... 60 °C
Set column oven temperature.
Autozero
Trigger analog output signal auto
zero.
DC_E -2 ... +2 V
Set
measuring potential (DC).
Filter
0.1, 0.2, 0.5 ... 5 s
Set electronic damping for the
analog output signal.
Scan_E1
-2 ... +2 V
Set sweep start potential E1.
Scan_E2
-2 ... +2 V
Set sweep end potential E2.
Scan_Cycle half,
full
Select scan cycle type.
Scan_Rate
1, 2, 5, … 50 mV/s
Set scan rate.
Links
The
Links
tab of the
817 Bioscan
is used for COM port selec-
tion and settings (details see Section 5.2.4 in the Software
Manual for IC Net 2.1).