4 opening criterion cassettes and removing the gel – Bio-Rad Criterion Dodeca Cell User Manual

Page 8

Advertising
background image

Table 4. Running Conditions (for 12 Criterion precast gels, 1.0
mm thickness)

Gel and Buffer

Voltage

Initial Current

Final Current

Run Time

System

(approximate)

(approximate)

(approximate)

Tris-HCl

200 V

1.0–1.4 A

400–500 mA

56 minutes

Tris/Glycine/SDS
or Tris/Glycine

The PowerPac 200 power supply will be limited at maximum power of 200W for a few minutes
(5–10 minutes) at this time the voltage will read less than the indicated voltage setting (about 180 V).
The Bromophenol Blue dye was run to the bottom of the gel.

Tris-Tricine

125 V

1.6 A

850 mA

110 minutes

Tris/Tricine/SDS

The PowerPac 200 power supply will be limited at maximum power of 200W for a few minutes
(20 minutes) at this time the voltage will read less than the indicated voltage setting (about 115 V).

TBE

200 V

500 mA

400 mA

76 minutes

Tris/Boric
Acid/EDTA

The gels were run until the first dye (Bromophenol Blue) cleared the gel.

TBE-Urea

200 V

500 mA

250 mA

76 minutes

Tris/Boric
Acid/EDTA

The gels were run until the second dye (Xylene Cyanole) was at the very bottom of the gel, the first
dye (Bromophenol Blue) was run off the gel

Note: Precast and Tris-Tricine gels will run slightly warmer then handcast, Tris-HCl and
TBE gels. Precast gels are prepared with Tris-HCl pH 8.6 and most handcast gels are cast
using Tris-HCl pH 8.8, this minor variation will alter the run times slightly, precast gels will
run a little faster.

2.4 Opening Criterion Cassettes and Removing the Gel

a.

After electrophoresis is complete, turn off the power supply and disconnect the electrical
leads.

b. Remove the cell from the stir plate. Be sure to pick up the tank from the bottom. Remove

the lid from the tank.

c.

Remove the gel cassette. Pour off and discard the buffer from the Integral Buffer Chamber.

Note: If the lower buffer will be reused, do not pour the depleted buffer from the integral
upper buffer chamber into the tank. The lower buffer can be reused up to 10 times depending
on usage (see guidelines in Section 3).