Pressure stability, Temperature stability, Sterilization – Bio-Rad Glass Econo-Column® Accessories User Manual

Page 5: Column cleaning, Accessories, Flow adaptors

Advertising
background image

Econo-Column

Flow Adaptor

Reagent

20°C

50°C

20°C

50°C

Citric acid

E

E

E

E

Cresol

G

F

G

F

Cyclohexane

F

N

F

N

Cyclohexanone

F

N

F

N

Dimethyl formamide

N

N

N

N

Dimethylsulfoxide

-

-

-

-

Ethyl acetate

N

N

N

N

Ethyl alcohol

E

G

E

G

Ethyl benzene

-

-

-

-

Ethyl chloride, liquid

F

N

F

N

Formaldehyde, 40%

E

G

E

G

Formic acid, 98–100%

E

G

E

G

Glycerine

E

E

E

E

Hydrochloric acid, 35%

E

G

E

G

Hydrofluoric acid, 48%

E

E

E

E

Hydrogen peroxide, 90%

E

-

E

-

Isobutyl alcohol

E

E

E

E

Isopropyl alcohol

E

E

E

E

Lactic acid, 85%

G

F

G

F

Methylene chloride

N

N

N

N

Nitric acid, 10%

E

E

E

E

Nitric acid, 50%

F

N

F

N

Nitrobenzene

N

N

N

N

Perchloric acid

G

N

G

N

Phenol, liquid

N

N

N

N

Phosphoric acid, 5%

E

E

E

E

Phosphoric acid, 85%

E

G

E

G

Propionic acid

E

G

E

G

Sodium acetate, saturated

E

E

E

E

Sodium hydroxide, 1%

E

E

E

E

Sodium hydroxide, 50% to saturated

E

G

E

G

Sodium hypochlorite

G

F

G

F

Sulfuric acid, 6%

E

E

E

E

Sulfuric acid, 60%

E

G

E

G

Sulfuric acid, 98%

F

N

F

N

Toluene

G

F

G

F

Trichloroacetic acid

F

N

F

N

Trichloroethane

N

N

N

N

Trichloroethylene

N

N

N

N

Tris buffer, solution

E

G

E

G

Urea

E

E

E

E

Xylene

F

N

F

N

Reprinted by permission of Nalge Company

2

LIT80d.qxd 6/11/2003 1:47 PM Page 2

Pressure Stability

Econo-Column chromatography columns are designed for low-pressure applica-
tions, i.e., less than 1.0 atm. Use caution when operating the column with a
pump. A disruption in flow may cause a rapid rise in backpressure. If this occurs,
immediately shut off the pump and check the gel bed for compression.

Temperature Stability

Econo-Column chromatography columns and flow adaptors can be operated at
temperatures up to 50°C. For applications requiring temperature control, jacket-
ed columns are available in various sizes and formats. Refer to the Bio-Rad Life
Sciences Research Products catalog for more information.

Sterilization

Econo-Column chromatography columns and accessories can be sterilized with
2 N NaOH or 100% ethanol. The columns can also be autoclaved, wet or dry, at
a maximum of 15 min at 121°C.

Column Cleaning

Econo-Column chromatography columns can be cleaned with 70% ethanol. The
bed support pore size is 20 µm and may require intermittent cleaning. To clean
the bed support, backwash with 1 N NaOH.

Accessories

A full range of Econo-Column accessories are also available. For more detailed
ordering information, refer to the Bio-Rad Life Science Research Products cata-
log.

Flow Adaptors

Flow adaptors improve resolution and reduce mechanical disruption by deliver-
ing sample and buffer directly to the top of the gel bed.

Catalog#

Description

738-0014

Flow Adaptor, 1 cm ID, 7 cm long

738-0015

Flow Adaptor, 1 cm ID, 14 cm long

738-0016

Flow Adaptor, 1.5 cm ID, 14 cm long

738-0017

Flow Adaptor, 2.5 cm ID, 14 cm long

738-0018

Flow Adaptor, 5.0 cm ID, 14 cm long

738-0019

Econo-Pac

®

Flow Adaptor, 1.5 cm ID, 7 cm long

3

LIT80d.qxd 6/11/2003 1:47 PM Page 3