ITC VerifyNow System Platelet Reactivity Test User Manual
Page 52

Retrieving Test Results - 7
Page 48
14439.C 03/2013
VerifyNow System User Manual |
USA
7.2.2
Wet Quality Control Results Log
The cursor is positioned on the most recent test result (Figure 7-5). Navigate up and down an individual
list using the ( , ) keys.
t
t
t
t
t
t
t
t
26-Feb-2012 13:51:12
WQC Results
09:31:25 S/N 0031
238 04061021
248 04061026
240 04061034
233 04061043
223 04061047
250 05021042
239 05021046
195 05021051
Rslt OpID MMDDHHMM
ALL
L
L
Figure 7‑5 WQC Results Log
WQC Result
Test Type
Operator ID
Date and Time
Print Result
Back
Print All Results
Feature
Description
WQC Result
Displays the numerical result for each WQC test performed.
Test Type
Displays a symbol indicating the type of WQC test performed.
IIb/IIIa Test (PAU resul
t)
Aspirin Test (ARU resul
t)
PRUTest (PRU resul
t)
Operator ID
Displays the operator who performed the test (when configured).
Date and Time
Stamp
Displays the date and time by month, day, hour (24 hr clock
format), and minute using the following notation, “MMDDHHMM.”
In addition to viewing earlier results, icons keys enable results to be printed (optional).
Icon
Description
Print Result
Scroll to the selected record, and select this icon to print the
selected WQC results.
Print All
Results
Scroll backwards through the results using the key, and
select this icon to print all WQC results in the log that are
forward of the cursor position
.
Back
Select this icon to return to the previous screen
.
7.3
View Instrument Usage Log
The instrument usage log shows how many tests of each type have been run, as well as the running
total of errors and alarms that have occurred during use. Use the following procedure to view the
Instrument Usage Log (Figure 7-6).
1.
From the main menu, enter
Operator ID and
Password, if required. Use the keypad to enter
the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to
enter the password number.
2.
Press the Maintenance
key.