WITHDRAWING GAS FROM THE CONTAINER

To withdraw gas from the XL-45HP/50HP/55HP/50VHP connect a suitable pressure regu-

lator to the USE connection, and the output of the regulator to your external equipment.

Then open the USE connection, and the output of the regulator to your external equip-

ment. Then open the USE and the PRESSURE BUILDING valves. When the container

pressure reaches 125 psig (8.6 bar/862 kPa), 300 psig (20.7 bar/2068 kPa) or 400 psig

(28 bar/2758 kPa) if equipped with the higher valve Dual Pressure Building/Economizer

regulator – set the pressure regulator for the desired delivery pressure.

Increasing Gas Supply Capacity – Two or more liquid containers may be manifolded

together. Accessory manifolds are available for use in creating a higher capacity gas

supply system. The XL-45HP/50HP/55HP/50VHP can supply gas at flow rates

1

up to

350 cfh @ NTP (9.2 cu. m/h @ STP) using only its internal vaporizer. At low flow rates, the

gas supplied will be at near ambient temperature. As the flow demand is increased, the

gas will become proportionately colder. If greater vaporizing capacity is required, an ac-

cessory external vaporizer is available. When an external vaporizer is sued, it must be

connected to the USE valve and the regulator moved to the output of the external vapor-

izer.

WITHDRAWING LIQUID FROM THE CONTAINER

When a container is used to supply liquid product, such as in an application as portable

distribution container for carbon dioxide, liquid may be withdrawn from the XL-45HP/50HP/

55HP/50VHP.

Attach a transfer hose to the LIQUID connection and open the adjacent LIQUID valve. The

presser in the container will drive liquid product out through the valve as the container

pressure exceeds that of the receiver.

The rate of liquid withdrawal from these containers is variable depending on the gas phase

pressure and the saturation temperature of the liquid.

1

See Specifications for your application.

CAUTION:

Internal orifices in

pressure regulators

used with CO

2

are

subject to the

formation of dry ice if

excessively cold gas

or high flow rates are

used. If this condition

occurs, an external

vaporizer should be

used to ensure the gas

is warmed before it

reaches the regulator.

Vaporizer Performance Graph

Liquid Withdrawal

CAUTION:

To avoid

contamination, close

the LIQUID valve on

an empty container

before disconnecting

the transfer line.