The Diagnostic (DIAG) mode menu allows
testing of specific areas of the scale’s function
and viewing current configuration settings. This
includes the following:

DISPLAY (DISP) – Shows the version and
revision of the software, followed by a display
segment test.

RAM (RA) – Performs a nondestructive test of
RAM in the processor.

ROM (RO) – Performs a checksum of all loca-
tions of ROM in the processor.

INPUT/OUTPUT (I/O) – Data is output by the
scale and through the use of a loopback con-
nector. The data is immediately read back into
the receive channel and verified against what
was sent. Requires a jumper (short) between
transmit and receive data lines.

DIVISION TEST, w/AZT (DIV-A) – Weight data
is normalized to 100,000 counts of displayed
resolution. AZT (Auto Zero Tracking) is enabled.

DIVISION TEST, w/o AZT (DIV-N) – Weight
data is normalized to 100,000 counts of dis-
played resolution. AZT is disabled.

Viewable configuration settings include:

LN FR

Line frequency

FILT

Filter setting

PROT

Serial protocol

BAUD

Baud rate

CAP

Scale capacity

UNITS

Unit of measure

Diagnostics Mode

Diagnostic (DIAG) Mode

Tip: Easy access to
the diagnostic mode is
available directly from
the front panel without
opening the scale to
access the Menu
mode switches. See
Setup Review/Test
section.