0 calibration, 1 gravity compensation, 2 front panel calibration – Rice Lake 920i Installation Manual V1.05 User Manual
Page 44: 1 gravity compensation 4.2 front panel calibration
40
920i
Installation Manual
4.0
Calibration
The
920i
can be calibrated using the front panel, serial commands, or
iRev
. Each method consists of the
following steps:
¥
Zero calibration
¥
Entering the test weight value
¥
Span calibration
¥
Optional Þve-point linearization
¥
Optional rezero calibration for test weights using hooks or chains
The following sections describe the calibration procedure for each of the calibration methods.
Figure 4-1. Calibration (CALIBR) Submenu
4.1
Gravity Compensation
Gravity compensation for latitude and elevation is available for the
920i
. To calibrate with gravity compensation,
the LOCALE parameter under the FEATURE menu must be set ON, and the LATUDE (latitude) and ELEVAT
(elevation, in meters) parameters set before calibrating the indicator (see Figure 3-8 on page 32).
If the indicator is later installed at a different location, gravity compensation can be applied to a pre-calibrated
indicator by adjusting the LATUDE and ELEVAT parameters.
4.2
Front Panel Calibration
The CALIBR submenu (under the SCALES menu,
see Figure 4-2) is used to calibrate the
920i
. The zero,
span, and linear calibration point displays provide a
set of softkeys used speciÞcally for calibration
procedures:
Default
Restores the value to its default
+/–
Toggles to allow entry of negative or
positive values
Calibrate
Performs calibration for the selected
point
Millivolts
(or
Counts
)
Toggles between display of captured
A/D counts and captured millivolts
values; allows entry of calibration
values in mV or counts
Restore
Restores the value to the previously
saved value
Figure 4-2. WZERO Calibration Display
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
ALGOUT
VERS
SCALE 1
CALIBR
WZERO
WVAL
WSPAN
WLIN
REZERO
PT->4
PT->5
PT->1
PT->2
PT->3