Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.
Display Cause
Solution
No segments appear on
No AC power is available
Check the AC power supply
the display
The power supply is not
Plug in the power supply
plugged in
high
The load exceeds the
Unload the balance
balance capacity
low or Err 54
Something is touching the
Move the object that is touching
weighing pan
the weighing pan
App.err.
Cannot store data:
Increase load
Load on weighing pan too
light or no sample on pan
while application is active
dis.err.
Display error:
Change the configuration in the
Data output not compatible
operating menu
with output format
prt.err.
Interface port for printer
Reset the menu factory settings
output is blocked
or
Contact your local
Sartorius Service Center
err 02
Calibration parameter not met, Calibrate only when zero is displayed
e.g.:
– Press w to tare the balance
– load on weighing pan
Unload the balance
err 10 The
w
key is blocked
After the tare memory has been
for active application programs; deleted using the c key,
Only 1 tare function can be
the w key can be used again
used at a time
err 11
Tara memory not allowed
Press w
The weight readout
Unstable ambient conditions
Set up balance in another area
changes constantly
(excessive vibration or draft)
Adjust Setup configuration
A foreign object is caught
Remove the foreign object
between weighing pan
and housing
The weight readout
The balance was not
Calibrated/adjust the balance
is obviously wrong
calibrated/adjusted
Balance not tared before
weighing
If any other errors occur, contact your local Sartorius Service Center.