Directions for Use

Printing & trends

69

Print OxyCRG

The OxyCRG is a printout of two minutes of continuous HR/PR and SpO

2

numerics, and a

compressed respiratory waveform.

1.

To print OxyCRG, from the Main Menu press SETUP, TRENDS, PRINT OXYCRG.

If any of the parameters have been completely inactive for the two minutes prior to
the printout, the associated band is empty.

If an SpO

2

or HR/PR alarm occurs, the OxyCRG prints 60 seconds later.

If an Apnea or RR/BR alarm occurs, the OxyCRG prints 75 seconds later.
Highlighted labels in the printout indicate which alarms occurred.

AUTO
TREND

Automatically print trends at the selected hours. Only the parameters
set to ON (for NIBP, RESP, P1, P2, or TEMP) are included in trend
printouts.