Directions for Use

Chapter 3 Calibrating the Spirometer

25

Figure 10. Calibration Code on Flow Transducer Package Label

3.

Press the desired softkey, as listed here.

Verify Calibration

To verify the accuracy of the system (without recalibrating). None of the
calibration factors will be saved. The date that prints on reports will be the last
calibration date, not the verification date.

Calibrate 1 Flow

To calibrate the system using one inhale/exhale cycle.

Calibrate 3 Flows

To calibrate the system using three inhale/exhale cycles at three different rates.

The “attach flow transducer” prompt appears, as shown in

Figure 11

.

Figure 11. “Attach Flow Transducer” Prompt

Drawing No. 30015257 VER. F

8

LOT

2

QTY 100

Disposable Flow Transducers

CALIBRATION CODE WKKVDXPB7

0050

200x-xx

4341 State Street Road
Skaneateles Falls, NY 13153 USA
www. welchallyn.com

REF 703419

Transducer
“cal” code

Lot code

Spirometer Calibration

9:17AM Oct 16 08

Attach the flow transducer

to the syringe,

Pull the plunger out,

Then select continue

Back

Continue

At any time, you can press Back to return to
the initial calibration screen, as shown in

Figure 9

on page 24.