Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 10

Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
8
37 Under minimum battery
volts.
Power/ Host
controller alert
60
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
38 Battery volts not set.
Module/ Host
controller alert
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
39 Predict algorithm not
set.
Module/ Host
controller alert
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
40 Ambient temperature
not set.
Module/ Host
controller alert
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
41 Non-responsive probe.
Thermistor pulled away
from the tip or heater
broken.
Probe/ Equipment
alert
32/33
E0.5/E0.9 Probe warmer
is not
responding.
Try another
predict cycle.
If the problem
persists,
replace the
probe.
Indicate
probe
error
8
42 Bad probe gain.
Probe/ Equipment
alert
34
E0.9
The probe was
not
characterized/
calibrated.
Calibrate or
replace the
probe.
Indicate
probe
error
8
43 Bad probe response
value.
Probe/ Equipment
alert
34
E0.9
The probe was
not
characterized/
calibrated.
Indicate
probe
error