Brookfield CAP1000 Viscometer User Manual

Page 17

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 

Manual No. M02-32A0607

Table III-2: Full Scale Range Viscosity for High Torque CAP 1000+

Cone

Number

Cone

Range

Constant

shear Rate

Constant

FsR Poise

at 00

RPM

FsR

Poise at

70 RPM

FsR Poise

at 900

RPM

Poise at any RPM

01

1875

13.33N

18.75

2.50

2.08

1875/N

02

3750

13.33N

37.50

5.00

4.17

3750/N

03

7500

13.33N

75.00

10.00

8.33

7500/N

04

15000

3.33N

150.00

20.00

16.67

1500/N

05

30000

3.33N

300.00

40.00

33.33

30000/N

06

75000

3.33N

750.00

100.00

83.33

75000/N

07

3150

2N

*

*

*

3150/N

08

12500

2N

*

*

*

12500/N

09

50000

2N

*

*

*

50000/N

10

5000

5N

50.00

6.67

5.56

5000/N

N= RPM

Poise x 100 = centiPoise

*use of this cone at these RPM is not recommended

You can also determine FSR for a speed setting that was selected at the time of order that is not

in the above tables by doing a simple calculation:
• For Low Torque CAP instruments:

Cone range constant / (22.7 * N) = FSR (Poise) where N = RPM

• For High Torque CAP instruments:

Cone range constant / N = FSR (Poise) where N = RPM

The last column in the above tables shows this calculation.

III.2 Accuracy of Viscosity and temperature

The following tables indicate the accuracy of the viscosity measurement for the CAP 1000+

Viscometer using CAP spindles 01-10. This accuracy depends on both the rotational speed of the

cone and the percent of Full Scale Range (%FSR) in Poise at which the viscosity is measured.

Accuracy is stated in Poise (P) and is calculated as a % of the FSR viscosity.

To calculate accuracy:
• Determine FSR viscosity in Poise for the torque model, cone spindle, and speed of rotation,

using the information in Tables III-1 or III-2.

• Find the column that best defines the speed of rotation used for the measurement and your

reported % FSR from the measurement.

• Consult the table below to determine the accuracy of your measurement
• Multiply the accuracy by the FSR viscosity if you need your accuracy stated in Poise.