Brookfield CAP1000 Viscometer User Manual
Page 17
Brookfield Engineering Labs., Inc.
Page
Manual No. M02-32A0607
Table III-2: Full Scale Range Viscosity for High Torque CAP 1000+
Cone
Number
Cone
Range
Constant
shear Rate
Constant
FsR Poise
at 00
RPM
FsR
Poise at
70 RPM
FsR Poise
at 900
RPM
Poise at any RPM
01
1875
13.33N
18.75
2.50
2.08
1875/N
02
3750
13.33N
37.50
5.00
4.17
3750/N
03
7500
13.33N
75.00
10.00
8.33
7500/N
04
15000
3.33N
150.00
20.00
16.67
1500/N
05
30000
3.33N
300.00
40.00
33.33
30000/N
06
75000
3.33N
750.00
100.00
83.33
75000/N
07
3150
2N
*
*
*
3150/N
08
12500
2N
*
*
*
12500/N
09
50000
2N
*
*
*
50000/N
10
5000
5N
50.00
6.67
5.56
5000/N
N= RPM
Poise x 100 = centiPoise
*use of this cone at these RPM is not recommended
You can also determine FSR for a speed setting that was selected at the time of order that is not
in the above tables by doing a simple calculation:
• For Low Torque CAP instruments:
Cone range constant / (22.7 * N) = FSR (Poise) where N = RPM
• For High Torque CAP instruments:
Cone range constant / N = FSR (Poise) where N = RPM
The last column in the above tables shows this calculation.
III.2 Accuracy of Viscosity and temperature
The following tables indicate the accuracy of the viscosity measurement for the CAP 1000+
Viscometer using CAP spindles 01-10. This accuracy depends on both the rotational speed of the
cone and the percent of Full Scale Range (%FSR) in Poise at which the viscosity is measured.
Accuracy is stated in Poise (P) and is calculated as a % of the FSR viscosity.
To calculate accuracy:
• Determine FSR viscosity in Poise for the torque model, cone spindle, and speed of rotation,
using the information in Tables III-1 or III-2.
• Find the column that best defines the speed of rotation used for the measurement and your
reported % FSR from the measurement.
• Consult the table below to determine the accuracy of your measurement
• Multiply the accuracy by the FSR viscosity if you need your accuracy stated in Poise.