ALTANA BYK-mac i COLOR Short Instructions User Manual

English, Operation, How to calibrate

Advertising
background image

269 020 584 E 0806

Additives & Instruments

A member of

Measure what you see.

BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments

Please note detailed operating instructions in English on CD-ROM.

ENGLISH

Operation

mode wheel:

Switch on

press

Show menu

press

Select function

scroll

Activate function press

esc button:

Change between menu

and measurement mode

std button:

Short-cut to measure standard

Operate button:

Switch on, select and measure

How to Calibrate?

Black/white calibration

Select Setup – Calibrate – Black/White Calibration

Place unit on black trap and confirm OK

Place unit on white standard and confirm OK

Black/white calibration is requested every 3 months.

Cyan check
After a black and white calibration the instrument must be checked on the
Cyan standard.

Place unit on Cyan standard and confirm OK

The Cyan check is requested every 30 days. Make sure to perform it in
laboratory environment (Temperature 20 - 25°C, Humidity 30 - 60%).

operate

How to measure Differences?

Select a stored standard, use

„AUTO standard“ or measure a standard

Input standard name:

Change character ==> turn scroll wheel
Move to next digit ==> press scroll wheel
Move one digit back ==> press esc

Input number of readings

Place instrument on standard

Press operate to take a reading

Copy predefined settings for data display

Continue with sample readings

Note: All color data inclusive spectral data are
saved and transferred to software.

First page displays color

differences

Scroll to second page for sparkle

and graininess results*

* Only available with BYK-mac i

Advertising
This manual is related to the following products: