Eagle Microsystems LP4310 User Manual

Page 4

Advertising
background image

LP4310 IB.DOC
9/16/04

2. The hook-up cable to the read-out should not run close to other unshielded cables. Display

instability may result.

3. For best accuracy, a flat, level, and rigid surface is recommended to support the scale.

5.

The area should be accessible for periodic cleaning.

2.4 INSTALLING ( refer to LP4310 Outline & Interconnect Dwg. at the end of this manual)

1.

Clean the site area of dirt and debris.

2.

The scale should be oriented so that loading will be from the hinge side of the scale. Secure the 2
“hinge brackets” to the floor with four (4) appropriately selected 3/8” diameter anchor bolts.

3. With the bottom of the scale exposed, install the leveling foot in the load cell. Screw the foot all

the way untill it bottoms out on the load cell.

2.5 LEVELING

Fig. 2: Detail showing easy leveling adjustments

with scale in place. A screw driver is the only tool needed.

1. For best scale performance scale should be level within 1/2 degree.

.
2.6 HOOK-UP CABLE

Table 1 shows the wiring connections necessary to attach the platform to the read-out instrument. The
color code and function are as noted.

C O L O R

FUNCTION

B L A C K

– EXCITATION

W H I T E

– SIGNAL

G R E E N

+ SIGNAL

R E D

+ EXCITATION

Table 1

SECTION 3.0 TROUBLE SHOOTING

The following is a list of potential problems and likely cures.

SLOT TYPE
SCREW DRIVER

ADJUSTMENT
ACCESS HOLE

FOOT

Advertising