Pump down function

Continuous pumping with pressure and time settings

• Operation of a vacuum pump via in-line valve

• Operation of a vacuum pump without in-line valve with VMS (Vac-

uum Management System, see “Accessories”, pg. 64)

Preselections
+

Use the selection knob to select the parameters. All parameters can

be altered even while operation control is running.

+

Minimum: The controller switches the pump off or closes the in-line

valve once the preset value for ”Minimum” has been reached.

Minimum” is adjustable in a range of 1-795 Torr (1-1060 mbar) or can

be set to ”Off”.

A preset ”Duration” (process time) has no effect if the process is

stopped due to a preset ”Minimum” before ”Duration” is reached.

+

Delay: Delay” determines the time the coolant valve remains open af-

ter the process has been stopped. Determines also the time the pump

(only with VMS module and in-line valve) remains running after the

process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately - and that

a pump with VMS module and in-line valve is switched off immediately

- when the process stops.).

+

Duration: Duration” determines the total process time since control

start.

The process time is adjustable between 1-1440 minutes (24 h) or can

be set to ”Off”. ”Off” indicates that no endpoint for pump down is deter-

mined.

If a ”Duration” is preset, the controller switches off the pump when the

preset process time is reached, even if a preset ”Minimum” is still not

reached.

+

If neither ”Minimum” nor ”Duration” is preset, process control has to be

stopped by pressing the ”STOP” key.